A multiple-channel cochlear implant and wearable speech-processor: an audiological evaluation.
Tong YC, Clark GM, Dowell RC, Martin LF, Seligman PM, Patrick JF
Acta oto-laryngologica
- 온라인 게재일
- –
- 저널 발행일
- 1981년
- Vol.92 No.3-4, pp.193-8
- PubMed 등록일
- 1981년 9월 1일
이 논문은 아직 한국어 요약이 없습니다.
버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.
영문 초록
Stand audiological tests were administered to a totally deaf multiple-channel cochlear implant patient with a wearable speech-processor in a monitored sound field under the following conditions: a wearable unit activated alone (WA), lipreading with the wearable unit off (LA), and wearable unit activated in combination with lipreading (WL). Thresholds obtained for narrow-band noise signals indicated that the wearable unit allowed the patient to detect a variety of sounds at different frequencies. The results obtained in closed-set word tests and open-set word and sentence tests showed significant improvements in word and sentence scores from LA to WL. In the open-set (C.I.D.) sentence test, the patient scored 22% for LA and 76% for WL. The WL score of 76% correlates with a satisfactory performance in understanding connected speech. The patient also scored 30% correct in a test involving the recognition of environmental sounds.
인용: Tong YC, Clark GM, Dowell RC, Martin LF, Seligman PM, Patrick JF A multiple-channel cochlear implant and wearable speech-processor: an audiological evaluation. Acta Otolaryngol. 1981;92(3-4):193-8. doi:10.3109/00016488109133256