Comparison of promontory and cochlear implants within a single subject.
Edgerton BJ, House WF, Doyle KJ
The Annals of otology, rhinology & laryngology. Supplement
- 온라인 게재일
- –
- 저널 발행일
- 1982년
- Vol.91 No.2 Pt 3, pp.100-3
- PubMed 등록일
- 1982년 3월 1일
이 논문은 아직 한국어 요약이 없습니다.
버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.
영문 초록
Audiometric pure tone, speech, and psychoacoustic data were obtained from a patient who had worn a promontory electrode for two years. Subsequently the promontory implant was replaced by a scala tympani electrode (cochlear implant) and testing was repeated. Warbled tone thresholds ranged from 52-68 dB SPL for the octave frequencies 250-8000 Hz with the cochlear implant (CI), and from 65-74 dB SPL with the promontory electrode (PE). Speech discrimination scores were better with the CI than with the PE, although performance on an environmental sounds test was slightly better with the PE. Smaller intensity difference limens (DLs) were obtained with the CI (1 dB) than with the PE (about 4 dB). Frequency discrimination was poor with both electrode systems. The CI provided better frequency DLs at 250, 500, and 1000 Hz; the PE produced better DLs at 2000 Hz. When the CI was combined with a hearing aid worn on the better implanted ear, performance improved over that with either prosthesis alone.
인용: Edgerton BJ, House WF, Doyle KJ Comparison of promontory and cochlear implants within a single subject. Ann Otol Rhinol Laryngol Suppl. 1982;91(2 Pt 3):100-3