Efficiency of a vibrating cochlear implant array for acoustic stimulation.
Ghoncheh M, Schurzig D, Prenzler N, Maier H
Hearing research
- 온라인 게재일
- 2026년 11월 6일
- 저널 발행일
- 미발행 (선게재)
- PubMed 등록일
- 2026년 11월 16일
이 논문은 아직 한국어 요약이 없습니다.
버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.
영문 초록
INTRODUCTION: Patients with conductive or mixed hearing loss can benefit from a combination of acoustic or mechanical stimulation and a cochlear implant (CI). We experimentally investigated the feasibility of using the CI electrode array as an acousto-mechanical stimulator in human temporal bones (TBs).
METHODS: First, the displacement magnitude of the stapes in response to the acoustic stimulation in the external auditory canal was measured in temporal bone specimens (N = 10). The scala vestibuli (SV) intracochlear pressure (PSV) in response to the acoustic stimulation was measured. Next, a CI (FLEX20, MED-EL, Austria) electrode was inserted into the scala tympani (ST) through the round window, and its distal end was attached to a commercial piezoelectric transducer using a custom-made adapter. Finally, the electrode displacement and PSV was measured in response to mechanical stimulation of the inserted CI electrode. The equivalent sound pressure level (eq. dB SPL) at 1 µm electrode displacement was determined to evaluate the achieved output level of the vibrating CI.
RESULTS: The vibrating CI produced an average equivalent output level of 108 eq. dB SPL (LDisp) at low frequencies (125 - 500 Hz) for a 1 µm electrode displacement of the CI. Above 1 kHz, the output increased by ∼10 dB per octave, reaching 103-139 eq. dB SPL, with a variability of 8-15 dB (SD) across experiments and frequencies up to 10 kHz.
CONCLUSION: The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.
인용: Ghoncheh M, Schurzig D, Prenzler N, Maier H Efficiency of a vibrating cochlear implant array for acoustic stimulation. Hear Res. 2026;483:109863. doi:10.1016/j.heares.2026.109863