Evaluation of bone-conducted sound during ear surgery: an in vivo pilot study.
Gille K, Fleischer M, Neudert M, Zahnert T, Bornitz M
Biomedical engineering online
- 온라인 게재일
- 2026년 11월 13일
- 저널 발행일
- 2026년 11월 13일
- Vol.25 No.1
- PubMed 등록일
- 2026년 11월 13일
이 논문은 아직 한국어 요약이 없습니다.
버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.
영문 초록
The impact of noise induced by drilling on hearing during ear operations has been discussed several times in the past. However, direct measurements of the equivalent sound pressure level in patients undergoing ear surgery have not previously been reported. This pilot study presents an easy-to-use method based on a calibrated accelerometer to determine the equivalent sound pressure level potentially applicable in the clinical routine. During seven monitored cochlear implant surgeries, the equivalent A-weighted sound pressure level LAF ranged from 111 to 126 dB(A)eqSPL at its maximum, whereas C-weighted peak levels LCpeak showed peaks up to 140 dB(C)eqSPL. The recommended daily noise exposure limit was exceeded for three of the seven patients, suggesting the potential for drilling-induced temporal shifts in hearing threshold. Based on these findings, standardized noise monitoring is recommended as a quality assurance measure during otological surgery.
인용: Gille K, Fleischer M, Neudert M, Zahnert T, Bornitz M Evaluation of bone-conducted sound during ear surgery: an in vivo pilot study. Biomed Eng Online. 2026;25(1). doi:10.1186/s12938-026-01652-0