Comparative Long-Term Outcomes of Lateral Wall and Slim Modiolar Cochlear Implant Electrodes in Children With Normal Cochlear Anatomy.
Zawawi F, Bukhari AF, Garrada M
Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
- 온라인 게재일
- 2026년 9월 30일
- 저널 발행일
- 미발행 (선게재)
- PubMed 등록일
- 2026년 9월 30일
이 논문은 아직 한국어 요약이 없습니다.
버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.
영문 초록
BACKGROUND: Electrode design may influence neural stimulation efficiency, speech outcomes, and long-term stability. While both lateral wall and slim modiolar electrodes are widely used in pediatric cochlear implantation, comparative long-term outcome data in children with normal cochlear anatomy remain limited.
METHODS: This retrospective comparative study included children with bilateral prelingual severe-to-profound sensorineural hearing loss who underwent cochlear implantation with either a lateral wall electrode (CI522/622) or a slim modiolar electrode (CI532/632) at a tertiary academic center between 2017 and 2021. All children had normal cochlear anatomy and a minimum follow-up of 4 years. Demographic data, surgical details, intraoperative electrophysiologic measures, postoperative audiologic and speech outcomes, data logging, and radiologic assessments were analyzed.
RESULTS: A total of 161 ears were included (83 lateral wall, 78 slim modiolar). Mean age at implantation did not differ between groups. The slim modiolar group demonstrated significantly lower intraoperative NRT thresholds and impedance values, higher 4-year speech discrimination scores (75.5% vs. 68.8%; P =0.001), and greater mean daily device use (11.2 vs. 9.7 h; P =0.02). Electrode migration occurred exclusively in the lateral wall group (8.4%), while none was observed in the slim modiolar group. Suspected scalar translocation was uncommon overall but more frequent with lateral wall electrodes (4.8% vs. 1.2%; P =0.02).
CONCLUSION: In children with normal cochlear anatomy, slim modiolar electrodes were associated with superior long-term speech outcomes, lower stimulation requirements, greater device use, and improved intracochlear stability compared with lateral wall electrodes.
인용: Zawawi F, Bukhari AF, Garrada M Comparative Long-Term Outcomes of Lateral Wall and Slim Modiolar Cochlear Implant Electrodes in Children With Normal Cochlear Anatomy. Otol Neurotol. 2026. doi:10.1097/MAO.0000000000005088