전체 논문 목록

Intraoperative high-resolution CT reveals electrode misplacement undetected by skull X-ray following cochlear implantation for IP-III inner ear malformation.

Zhao Z, Guo J, Feng T, Huang Y, Liu W, Zhang Z, et al.

BMJ case reports

온라인 게재일
2026년 9월 16일
저널 발행일
2026년 9월 16일
Vol.19 No.9
PubMed 등록일
2026년 9월 16일

이 논문은 아직 한국어 요약이 없습니다.

버튼을 누르면 Gemini가 초록을 요약합니다 (20~40초). 만들어진 요약은 모든 방문자에게 보입니다.

영문 초록

Incomplete partition type III (IP-III) malformation presents unique challenges for cochlear implantation. We report a critical intraoperative imaging comparison in a paediatric IP-III case where traditional skull X-ray failed to reveal the precise location of the cochlear electrode, while high-resolution CT (HRCT) precisely identified the site of electrode displacement. This finding demonstrates a significant limitation of routine radiography in these complex anatomies. Our experience underscores HRCT's critical role in accurately verifying electrode placement for patients with IP-III. Recognising this imaging discrepancy is essential for preventing undiagnosed malposition, guiding surgical revision and optimising auditory outcomes. Our findings support considering intraoperative HRCT as a valuable adjunct for confirming electrode position in IP-III implantation. Further multicentre studies are warranted to establish evidence-based imaging protocols for this challenging patient group.

인용: Zhao Z, Guo J, Feng T, Huang Y, Liu W, Zhang Z, et al. Intraoperative high-resolution CT reveals electrode misplacement undetected by skull X-ray following cochlear implantation for IP-III inner ear malformation. BMJ Case Rep. 2026;19(9). doi:10.1136/bcr-2025-269858